What is the debate over the
Intoxilyzer's reliability and accuracy?
Proponents
of the Intoxilyzer say it will only show a result from absorbed
breath (deep lung air) alcohol and nothing else. However,
opponents say that the Intoxilyzer often misreads other commonly
found substances in human breath and erroneously gives high
readings saying that they are from alcohol.
Of
particular importance here are the following facts. First, the
DWI alcohol concentration law says a person is intoxicated when
he has a .08 concentration in his breath, but, it does not say
.08 by Intoxilyzer. This fact means that no judge or jury is
required to believe that an Intoxilyzer result of .08 or more is
accurate or reliable.
Second,
neither the manufacturer nor the DPS will allow anyone, other
than law enforcement personnel, to test either the machine's
accuracy or its reliability. It is generally understood that for
a procedure to be accepted as accurate and reliable in science,
that it must be open and available for the scientific community
to test and retest the procedure. This is not the case with the
Intoxilyzer.
Third, the
manufacturer says it does not warrant that the Intoxilyzer is
fit for any particular purpose. This fact clearly is an implicit
admission by the manufacturer that its machine is not even
warranted as accurate and reliable for breath testing.
Fourth, the
Intoxilyzer is capable of breath preservation, however, our DPS
purposely fails to require the breath specimens to be saved. The
cost of preservation would be less than $2.00 per test and would
allow an opportunity for the person charged with DWI to check
the accuracy of the sample. And, if found to be inaccurate,
attack the validity of the prosecutor's test. Indeed, it is a
generally accepted scientific fact that the re-testing of
preserved breath specimens, which is done by a method known as
gas chromatography, is a more accurate and reliable means of
breath alcohol concentration testing than that done by the
Intoxilyzer.
Fifth and
last, the Intoxilyzer's working design is premised on the
assumption that every person tested is exactly the average
person. All persons are not exactly average! Human beings
come in all different sizes, weights, ages, muscle tones, lung
capacities, alcohol tolerances, temperatures, hematocrit levels
(amount of solids in the blood) and blood/breath ratios (the
number of times an item appears in the blood vs. the number of
times the same item appears in the breath). Automatic and
undetected error can be illustrated by simply having the person
tested not be exactly average. In this regard, it should be
noted that the Intoxilyzer assumes a blood/breath ration of
2100/1 (i.e., 2100 parts of alcohol in the blood for every 1
part of alcohol in the breath) for every person tested. Here, it
can be noted that a majority of persons have a blood/breath
ration of 2100/1 or greater. Persons with a higher blood/breath
ration of 2100/1 will not be prejudiced by the Intoxilyzer's
assumption. However, persons with a lower blood/breath ration
will be prejudiced because the Intoxilyzer will erroneously read
too high of an alcohol concentration result, thus potentially
causing a person who should test at .04, .05, .06, etc. to
actually test out at .08, .11, .12, etc. Of particular import
here is the fact that scientists have documented persons with
blood/breath ratios as low as 1100/1.
This same
type of prejudice also occurs where the person tested is not
exactly average with respect to other bodily functions: muscle
development, temperature, hematocrit level, etc. Moreover, since
the machine was built by humans, is serviced by humans, and is
operated by humans, it is subject to human error just like all
other machines. The above facts conclusively demonstrate that
the Intoxilyzer, even if it is properly working and is being
properly operated, because the person being tested is not
exactly average, can label an innocent person as guilty.